Freon Gas Detector
OSC92GT200, 200A
<Features>
- Reduces reaction to adhesive agent and Styrofoam which are used for alcoholic solvent and PCB board
- Suitable for leak detect inspection at factories
- Environmental pollution level is displayed on LED screen in 10 steps colors
- Basic consumables are cartridge filter only
Only for OSC 92GT200A
- Calibration gas can be used for more than 10 year without replacement
MODEL | OSC 92GT100 | |||
Objective gas | Existing general fleon gases such as R12, R22, R502, R134a, R404A, R407C, R410A, R600a | |||
Detecting method | Semiconductor sensor suction method | |||
Maximum sensitivity | 5X10-6ml/sec (approx. 0.5g/year) in case of R134a | |||
Reaction speed | within 1.5 sec (at range H5, suction volume 250 mL/min) | |||
Sensitivity range selection | H/L ranges 5 steps each | |||
Suction method | Auto suction pump (max. 300 mL/min) | |||
Leak indication | Analog meter and buzzer (buzzer volume and alarm point adjustable) | |||
Calibration device (only OSC 92GT200A) | Incorporated NIST authorized calibration device R134a 1.1〜2.9 g/year | |||
Power source | AC100V-240V (50/60Hz) | |||
External output | Voltage output for recorder DC0-5V, Alarm contact output (A point AC125V 1A) | |||
Operational environment | 20±15℃ 65±20%RH | |||
Dimensions and weight | 350W x 270D x 160H mm, approx. 8.7 kg | |||
Standard accessories | Standard probe (2 m) x 1, Spare fuse (3A) x 1, Cartridge filter x 5, perfornace data sheet, operation manual, certificate of guarantee 2 | |||