Featured Functions

-Biaxial balance is adopted.
It can minimize inertia force and external forces generated by sliding motion.

-Measurements in both ways:
This mode obtains coefficient of friction data in both ways of stage travel. It can reduce time when measuring numbers of cycles.

-Function of lifting contact part can be selected when measuring with one way mode.
The contact part is lifted automatically in the return travel of stage.

-Repetitive static friction measurements:
This mode repeats move and stop of stage, and 12 data of coefficient of static friction are obtained in a stroke.

-Standard repeat measurements:
Both static and kinetic coefficients of frictions are measured repeatedly on same measuring area. (up to 12 cycles)

-Change sample measurements:
for comparison of data with different samples or different areas of a sample.

-Dependency on repeat:
continuous reciprocating measurements to evaluate change of coefficient of friction with times. The data can show wear of surface in the view of friction. (Max. 10,000 times)

-Dependency on weight:
for study how coefficient of friction changes by change of vertical loading

-Dependency on speed:
for study how coefficient of friction changes by changing moving speed

-Dependency on stop-time:
This shows the coefficient of friction changes with elapsed time of surface. The stage repeats moving and stopping for a specified time interval.



Specification
ModelOSC
92LJ
113A
OSC
92LJ
113B
OSC
92LJ
113C
OSC
92LJ
113D
Measuring methodBowden method
Display resolutionCoefficient of friction: 0.0001
Frictional force rangeStandard: 0-1,000gf (9.8N)
Option: 0-2,000gf (19.6N)d
Vertical
load
100-1,000g (in steps of 50g)[When using
option: 200-2,000g (in steps of 100g)]
Travel distance1 - 80mm (in steps of 1mm)|
Travel speed0.1 - 100.0mm/sec1 - 600rpm
Travel timesMax.
10,000
times
1-12
times
Max. 999,999 cycles
Max. sample size180(W)~100(D)
~30(H)mm
ƒΣ150x
10(H)mm
ƒΣ10-100
x300(L)mm
Stage size180(W)~100(D)mmƒΣ150mmDepends on
sample size
Temperature controlavailable at options Jacket stage: 10 - 60‹C Heater stage: room temp. - 180‹Cat room temperature
External dimensionsabout 600(D)~310(W)~420(H)mmabout 600(D)~
425(W)~
420(H)mm
Weightabout 29kgabout 32kgabout 29kg
Power supplyAC100?240V, 50/60Hz
Environment for useTemp.: 15-30‹C,
Humidity: 30-80%RH (non condensing)